RESPIRONICS, INC. DREAMSTATION BIPAP PRO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX600S11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles related to cpap device's sound abatement foam.There was no report of medical intervention.There was no report of serious or permanent harm or injury.The device was returned to the third-party service centre for further evaluation.During the evaluation of the device at the third-party service centre, the device was visually inspected and found no evidence of visible foam particles.The third-party service centre concludes that they couldn't confirm the customer's allegation and device was scrapped.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP PRO
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key21585468
Report Number2518422-2025-032865
Device Sequence Number8757752
Product Code BZD
UDI-Device Identifier00606959424234
UDI-Public(01)00606959424234
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDSX600S11
Device Catalogue NumberDSX600S11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Initial Report FDA Received Date03/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-