• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VIANT MEDICAL, LLC VIANT MEDICAL, LLC; OFFSET REAMER HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VIANT MEDICAL, LLC VIANT MEDICAL, LLC; OFFSET REAMER HANDLE Back to Search Results
Model Number T17653
Medical Device Problem Codes Break (1069); Unstable (1667); Material Integrity Problem (2978); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint sample was not returned to viant for evaluation.Thus, the reported event is non-verifiable.The current ifu sent with this device today, man-004006 rev.A, states the following; end of life is determined by wear and damage due to intended use, visually inspect for damage and wear.If the instrument is damaged and worn it is considered at the end of its life and should be discarded, where instruments form part of a larger assembly, check assembly with mating components, check hinged instruments for smooth movement, each component of the viant offset reamer handle is not a medical device, all components when assembled create the medical device.Therefore, the components of each viant reamer handle form a unique set when assembled.In no instance should the components from one offset reamer handle be mixed with components from another reamer handle.The unique device identification for this medical device is located on the assembled offset reamer handle, viant devices should only be used by qualified personnel fully trained in the use of the surgical instruments and the relevant surgical procedures, do not modify viant instruments in any way and handle with care at all times.Surface scratches can increase wear and the risk of corrosion, manual surgical instruments have a limited life-span which is determined by wear or damage due to repeated intended use.When a surgical instrument reaches the end of its functional life, clean the instrument of any and all biomaterial/biohazards and safely discard the instrument in accordance with applicable laws and regulations.The device history record (dhr) review was not performed as the lot number is unknown.The viant risk management files were reviewed to ensure the reported failure mode (or similar) is captured and assessed.The review revealed there are similar failure modes identified and mitigated to the lowest possible risk region.In conclusion, the complaint sample was not received by viant for evaluation and the reported event is non-verifiable.If the complaint sample is received by viant, it will be evaluated and the complaint record will be updated accordingly.No further investigation with regard to this complaint is required at this time.Viant will continue to monitor for trends.D9, h3: the distributor stated that all broken pieces were accounted for, removed from the field, and disposed of.The device will not be returned for evaluation.G2: complaint information provided by distributor, depuy synthes.
 
Event or Problem Description
It was reported during an unknown patient procedure that while reaming acetabulum the surgeon noticed that the reamer was not reaming symmetrically.The reamer handle was taken apart and the inner shaft of the reamer was broken.All broken pieces were accounted for, removed from the field, and disposed of.There was 5 minute surgical delay in surgery while staff got another tray.No consequences or impact to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIANT MEDICAL, LLC
Common Device Name
OFFSET REAMER HANDLE
Manufacturer (Section D)
VIANT MEDICAL, LLC
4545 kroemer road
fort wayne IN 46818
Manufacturer (Section G)
VIANT MEDICAL, LLC
4545 kroemer road
fort wayne IN 46818
Manufacturer Contact
tony singh
4545 kroemer road
fort wayne, IN 46818
MDR Report Key21585594
Report Number3004976965-2025-00003
Device Sequence Number21199209
Product Code LXH
UDI-Device Identifier00840096400083
UDI-Public00840096400083
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberT17653
Device Catalogue NumberT17653
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2025
Initial Report FDA Received Date03/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-