• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. CLOSURE TOP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. CLOSURE TOP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 07.02010.001
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D4: the remainder of the udi number is unknown because the lot number is not available.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2025-00079 through 3012447612-2025-00083.
 
Event or Problem Description
It was reported that two vital mis set screws migrated from their mating polyaxial pedicle screws post-operatively.This was discovered approximately two weeks post-op via x-ray.The patient is currently asymptomatic but the surgeon is planning to revise the construct.The surgeon believes that the torque handle used did not provide adequate torque.This is report three of five for this event.
 
Additional Manufacturer Narrative
H6: additional method code: 4110.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the device was not returned; however, x-ray images were provided which show that two closure tops have migrated out of their screws.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Dhr review: the lot number was not reported and the device was not returned, therefore a dhr review is unable to be performed.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that two vital mis set screws migrated from their mating polyaxial pedicle screws post-operatively.This was discovered approximately two weeks post-op via x-ray.The patient is currently asymptomatic but the surgeon is planning to revise the construct.The surgeon believes that the torque handle used did not provide adequate torque.This is report three of five for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSURE TOP
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key21586641
Report Number3012447612-2025-00081
Device Sequence Number15816867
Product Code NKB
UDI-Device Identifier00889024003170
UDI-Public(01)00889024003170
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K150896
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number07.02010.001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/12/2025
Supplement Report FDA Received Date08/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-