| Catalog Number |
07559674190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The c513 analyzer serial number was (b)(6).The field application specialist (fas) flushed the sample probe and the reagent probe.The fas checked the rinse station and replaced a sample probe.Qc results were not acceptable.The fas performed a mechanism check and observed water droplets on the cell rinse and cuvettes.The field service engineer (fse) replaced the other sample probe and other parts; the reagent probe was bent and was replaced.As several service actions were performed, the specific cause of the event could not be determined.The service actions resolved the issue.
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of discrepant high results for 2 patient samples tested for tina-quant hemoglobin a1cdx gen.3 (a1cdx) on a cobas c 513 analyzer.Patient 1 initial result was 7.11%.The repeat result from a different c513 analyzer was 5.34%.Patient 2 initial result was 14.58%.The repeat result from a different c513 analyzer was 5.97%.The repeat results were believed to be correct.
|
| |
|
Search Alerts/Recalls
|