• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH FEMORAL NAIL, A/R T2 FEMUR Ø10X360 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH FEMORAL NAIL, A/R T2 FEMUR Ø10X360 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 18251036S
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 09/24/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
As per the legal department'' bghm (berufsgenossenschaft holz und metall) insurance have reported that there is an increase in lead, chromium, cobalt, zinc and cadmium has been detected in their insured¿s blood and urine.An allegation has been made by the insured patient¿s treating physician that this complaint may be due to the t2 femoral nail implanted in the patient in 2007.".
 
Event or Problem Description
As per the legal department'' bghm (berufsgenossenschaft holz und metall) insurance have reported that there is an increase in lead, chromium, cobalt, zinc and cadmium has been detected in their insured¿s blood and urine.An allegation has been made by the insured patient¿s treating physician that this complaint may be due to the t2 femoral nail implanted in the patient in 2007.".
 
Additional Manufacturer Narrative
Please note the correction to d4 lot#.The reported event could not be confirmed during the investigation the reported event could not be confirmed since the device was not returned for evaluation and no other evidence were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.As requested, the chemical composition (mass percentage) per astm f136 of the implant's material is given below: carbon less than 0.08 aluminium 5.5 ¿ 6.5 vanadium 3.5 ¿ 4.5 iron less than 0.25 oxygen less than 0.13 nitrogen less than 0.05 hydrogen less than 0.012 titanium: balance a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL NAIL, A/R T2 FEMUR Ø10X360 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21587628
Report Number0008031020-2025-00402
Device Sequence Number9468013
Product Code HSB
UDI-Device Identifier04546540199515
UDI-Public04546540199515
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K010801
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/30/2008
Device Catalogue Number18251036S
Device Lot NumberK887190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2025
Supplement Date Received by Manufacturer04/01/2025
Initial Report FDA Received Date03/12/2025
Supplement Report FDA Received Date04/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2004
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-