| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6-investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a lens rotation.On separate visit the lens was repositioned, and problem did not resolve.On a third visit the lens was exchanged for a longer length lens and the problem was resolved.Cause reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3-device evaluation: lens returned in a microcentrifuge vial with moisture.Visual inspection found no visual damage to lens.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|