| Model Number |
TRIEEAXL28XT |
| Medical Device Problem Codes |
Failure to Align (2522); Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, during setup prior to firing, it was noted that the tip of the circular staler was bent.The anvil tilted prior to firing.The anvil would attach to the body of the eea stapler.Another device was used to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9, g3, pma / 510(k) #, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The visual inspection of the staple guide noted the presence of a full complement of staples.Further inspection noted that the green bar was not visible in the indicator window and the safety feature was engaged.The staple guide was observed to be attached to the instrument with no damage to the staple guide.The anvil of the instrument was observed to be not tilted and separated from the instrument.Further inspection of the anvil cutting ring showed no traces of knife impression and the ring was received intact.One of the retainer legs of the anvil was observed to be bent.Due to the observed retainer leg damage of the clinical anvil, a representative anvil was attached to the center rod without difficulty prior to firing the instrument.The wing nut and safety functioned properly upon rotating the twist knob.The green bar was visible within the indicator window when the twist knob was fully closed.The instrument was applied to the appropriate test media producing acceptable results.Pusher-fingers and knife advanced when the handle was squeezed and the knife cut the media cleanly and completely and a full complement of staples was deployed and properly formed.The device also produced all audible, tactile and visual indicators during the functional testing.The instrument body label was removed for evaluation of the eccentric washer assembly.Evaluation of the eccentric washer noted that it was secured.It was reported that the anvil was difficult to attach.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.Anvil damage may occur if the anvil was manipulated with excessive force during the attachment to the instrument, or if the anvil was mishandled during the removal of fitting accessories.It was also reported that the anvil tilted prematurely.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, during setup prior to firing, the anvil tilted.The anvil would not attach to the body of the stapler.Another device was used to complete the case.There was no patient injury.
|
| |
|