• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,MCOL MG,4.1MM,11.5MML; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,MCOL MG,4.1MM,11.5MML; DENTAL IMPLANT Back to Search Results
Catalog Number TSVM4B11
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight: unknown / not provided d4: additional device information: unknown / not provided.H4: device manufacturer date: unknown / not provided.
 
Event or Problem Description
The doctor reports that the dental implant located in position number #36 failed because it fractured at the head and was removed.The doctor reports that the procedure was not concluded by placing another implant.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) tsvm4b11, (imp,tsv,mcol mg,4.1mm,11.5mml) for evaluation.Visual evaluation was performed, the implant has signs of use, apparent bone / tissue attached to external threads.The implant was observed fractured at the collar region and a fractured screw fragment was seen stuck inside the implant.Device history record (dhr) review was completed for the subject lot number 1257136.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1257136 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : implant.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did occur.The implant was fractured at the collar.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,MCOL MG,4.1MM,11.5MML
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21591650
Report Number0002023141-2025-00644
Device Sequence Number18374045
Product Code DZE
UDI-Device Identifier00889024019638
UDI-Public(01)00889024019638(17)270804(10)1257136
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K111889
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberTSVM4B11
Device Lot Number1257136
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/17/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-