During an unknown endoscopic procedure, the physician used a cook acusnare polypectomy snare soft.It was reported [that] when capturing the polyp and proceeding with the cut, the plastic part wrinkled and no current passed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation evaluation: a product evaluation was performed only by the picture provided in response to this report because the product said to be involved was not provided to cook for evaluation.Per the photo provided we cannot complete a full evaluation.Without the product or substantial evidence to contradict the complaint, it is considered confirmed based solely on statements and photos describing the event.The picture provided shows the pouch with the device in the coiled position.Local labeling can be seen on the pouch.No conclusions can be made about the device as the base of the handle at the sheath cannot be seen.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the photo could not confirm the report.We could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." instructions for use states: "fully retract and extend snare to confirm smooth operation of device." instructions for use states: "inspect active cord.Cord must be free of kinks, bends, breaks, and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|