| Model Number |
GALILEO® CAPTURING ROD |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 02/25/2025, it was reported by a sales representative via (b)(4) that a 5046-100 lag screw capturing rod nut and a 5033-100 capturing rod assembly had an issue with not locking the lag screw into the nail as designed and had to use the locking end cap instead.This was discovered during a procedure, with no reported patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpackaged 5033-100 capturing rod batch number: 220097 was received for investigation.Visual evaluation of the returned device noted that a 5046-100 lag screw capturing rod nut had been improperly threaded onto the returned 5033-100 capturing rod and the devices could not be separated.It was further noted that the threads where the lag screw mates were also damaged.Functional testing was not able to be performed due to the damage to the device.The most likely cause for the reported failure can be attributed to misuse due to use error when threading the mating devices.Refer to investigation photos.
|
| |
|