• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. GALILEO® CAPTURING ROD; INTRAMED FIXATN ROD AND ACCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. GALILEO® CAPTURING ROD; INTRAMED FIXATN ROD AND ACCESS Back to Search Results
Model Number GALILEO® CAPTURING ROD
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 02/25/2025, it was reported by a sales representative via (b)(4) that a 5046-100 lag screw capturing rod nut and a 5033-100 capturing rod assembly had an issue with not locking the lag screw into the nail as designed and had to use the locking end cap instead.This was discovered during a procedure, with no reported patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Complaint allegation is confirmed.One unpackaged 5033-100 capturing rod batch number: 220097 was received for investigation.Visual evaluation of the returned device noted that a 5046-100 lag screw capturing rod nut had been improperly threaded onto the returned 5033-100 capturing rod and the devices could not be separated.It was further noted that the threads where the lag screw mates were also damaged.Functional testing was not able to be performed due to the damage to the device.The most likely cause for the reported failure can be attributed to misuse due to use error when threading the mating devices.Refer to investigation photos.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALILEO® CAPTURING ROD
Common Device Name
INTRAMED FIXATN ROD AND ACCESS
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21593298
Report Number1220246-2025-00911
Device Sequence Number18348660
Product Code HSB
UDI-Device Identifier00848665039177
UDI-Public(01)00848665039177(11)220523(10)220097
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K202099
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGALILEO® CAPTURING ROD
Device Catalogue Number5033-100
Device Lot Number220097
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2025
Supplement Date Received by Manufacturer06/10/2025
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date07/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-