Model Number DSX400H11C
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.No other clinical information or medical interventions were reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of particles.The device was returned to the third-party service centre for further evaluation.During the evaluation of the device at the third-party service centre, the device was visually inspected and found no evidence of visible foam particles.The technician confirmed no particles in the air path.During the evaluation, no secondary findings was observed.There was one error code found.The third-party service centre concludes that they couldn't confirm the customer's allegation and unit scrapped.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.In box h: evaluation method code, evaluation result code and conclusion code has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION CPAP PRO
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key21593309
Report Number2518422-2025-033754
Device Sequence Number18377389
Product Code BZD
UDI-Device Identifier00606959025608
UDI-Public(01)00606959025608
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDSX400H11C
Device Catalogue NumberDSX400H11C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer05/13/2025
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date08/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/18/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-