| Catalog Number |
0168L12 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the foley catheter after retention, a natural removal event had occurred, and there was a possibility that sterilized water was leaking from the balloon part but specific part was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter after retention, a natural removal event had occurred, and there was a possibility that sterilized water was leaking from the balloon part but specific part was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is unconfirmed.Photo: received three (3) photo samples.All photo samples showcase an overview of foley catheter.Visual: received one (1) used foley catheter without original packaging.Visual inspection noted no obvious observations.The catheter balloon was inflated with 10ml methylene blue solution (3 drops 1 percentage aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated in 11 seconds with no cuffing or leaks noted, returning 10ml of solution.This is within specification which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.No root cause could be found because the reported event was unconfirmed.A review of the device history record was not necessary due to the reported event being unconfirmed.As the reported event is unconfirmed, a risk and labeling review is not required.Correction- d, e, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|