| Model Number |
PXAVMP |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been discarded at the facility and will not be returned for product evaluation.The device history record review is pending, once the results are available they will be submitted in a supplemental report.Additional product code dxo.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a failure with a pressure monitoring kit.The tubing on the device broke and became disconnected on the patient side of the flow restrictor.This occurred during patient use.There was no patient harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
There is no product evaluation results as the device was discarded and not returned.Without its return, it is not possible to determine if there was damage or a defect that existed on the unit that could have contributed to the event.It is not known if any procedural factors may have contributed to the reported issue.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.As the complaint could not be confirmed, there is not sufficient evidence to determine a root cause.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as part of this monthly review.The device history record review was performed and all manufacturing inspections passed with no non conformances.
|
| |
|
Search Alerts/Recalls
|