Additional information provided in d.9., h.3., h.6.And h.11.One opened probe, with tip protector, in a pouch was received for the report.The returned sample was visually inspected and was found to be conforming.The sample was then functionally tested for actuation and aspiration and was found to be conforming for both functional tests.A review of the device history record traceable to the lot number obtained from the device's radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be conforming, therefore an aspiration failure as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the returned probe was manufactured to specifications.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|