Catalog Number 8065751561
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that the anterior vitrectomy cutter could not aspirate during cataract surgery (with intraocular lenses implant).The condition of actuation and cutting was unknown.The product was replaced with another one with the same lot number, and the surgery was completed.There was no patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.One opened probe, with tip protector, in a pouch was received for the report.The returned sample was visually inspected and was found to be conforming.The sample was then functionally tested for actuation and aspiration and was found to be conforming for both functional tests.A review of the device history record traceable to the lot number obtained from the device's radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be conforming, therefore an aspiration failure as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the returned probe was manufactured to specifications.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION ULTRAVIT PROBE PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21596470
Report Number1644019-2025-00669
Device Sequence Number15688165
Product Code LRO
UDI-Device Identifier00380657515615
UDI-Public(01)00380657515615(17)250930(11)231004(10)1680YP
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue Number8065751561
Device Lot Number1680YP
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/18/2025
Supplement Date Received by Manufacturer05/31/2025
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-