• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ILS 33MM, CURVED; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ILS 33MM, CURVED; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECS33A
Medical Device Problem Code Ejection Problem (4009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 3/13/2025.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a colectomy surgery, noted the anvil lock-out loose.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent 4/11/2025 the device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Additional Manufacturer Narrative
(b)(4).Date sent 4/21/2025.D4 batch #u5c30d.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ecs33a device arrived with no apparent damage.The breakaway washer uncut and indented and there were 10 partially formed staples and protruding on the pockets.The protruding staples were removed, the device was reloaded with staples and tested for functionality with a test washer; the device formed all the staples, as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.The condition of the washer indicates that the device had not been fired through a full firing stroke or possibly that the orange indicator was not fully into the safe green firing range.Before firing the device the orange indicator should be fully within the green range of the gap setting scale and the safety should not be released prior to firing.Also, if the firing sequence is not fully completed (the firing handle reaches its stopping point, and the firing trigger is parallel to the instrument handle) staples could be partially deployed without cutting the washer.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number u5c30d, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILS 33MM, CURVED
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key21596903
Report Number3005075853-2025-01973
Device Sequence Number13020370
Product Code GDW
UDI-Device Identifier10705036012412
UDI-Public(01)10705036012412(17)250228(10)U93V7F
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K940967
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue NumberECS33A
Device Lot NumberU93V7F
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2025
Supplement Date Received by Manufacturer04/07/2025
04/11/2025
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date04/11/2025
04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-