• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALPHATEC SPINE, INC. ARSENAL SPINAL FIXATION SYSTEM; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALPHATEC SPINE, INC. ARSENAL SPINAL FIXATION SYSTEM; SCREWDRIVER Back to Search Results
Model Number 87134
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual inspection confirms that the tip of the driver is twisted and chipped.The root cause is likely due to wear as a result of improper loading before torque delivery through the tip.Review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation and/or breakage of device components.
 
Event or Problem Description
It was reported that the driver twisted while final tightening a screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARSENAL SPINAL FIXATION SYSTEM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer Contact
wesley channell
1950 camino vida roble
carlsbad, CA 92008
9014283693
MDR Report Key21596969
Report Number2027467-2025-00046
Device Sequence Number8939206
Product Code HXX
UDI-Device Identifier00840967197906
UDI-Public(01)00840967197906(10)7897401
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number87134
Device Catalogue Number87134
Device Lot Number7897401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2025
Initial Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-