| Model Number |
MR8-AA10-R |
| Medical Device Problem Codes |
Corroded (1131); Detachment of Device or Device Component (2907); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that device had distal tube bearing is detached.The likely cause of failure is corrosion.It was also found that input and output shaft bearings are corroded, spacer and shaft is worn, and other tube bearings are corroded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of bearing detachment and corrosion and worn.No patient impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11 : based on device evaluation and additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned or met reporting requirement criteria.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|