C.R. BARD, INC. (BASD) -3006260740 SAFESTEP; PORT ACCESS NEEDLE Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Twisted/Bent (2981); Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
 
Event or Problem Description
It was reported that the safety feature did not work.Patient came to the hospital and attempted to remove the needle, but the safety mechanism failed to activate and the needle was removed.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of difficulty advancing the safety mechanism was confirmed and determined to be caused by a bent needle.However, the exact cause of the bent needle remains unknown.The product returned for evaluation was a one 22g safestep infusion set w/ y-site.The returned product sample was evaluated and the needle was observed to be bent near the needle base.The following observations were noted during the sample evaluation: ¿ usage residue was seen on the sample which proved that the product had experienced at least some use ¿ the tip of the needle bevel appeared unremarkable an attempt to advance the safety was successful; however, resistance was encountered when passing the safety over the bent region of the needle.Based on the evidence provided with the returned sample it is unknown when or how the bend occurred in the needle.Damage to the needle could have occurred at the manufacturing facility, during shipping, during use, or during storage of the product, and no evidence was observed which supported a specific root cause of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP
Common Device Name
PORT ACCESS NEEDLE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key21600541
Report Number3006260740-2025-01473
Device Sequence Number8939351
Product Code FPA
UDI-Device Identifier00801741066177
UDI-Public(01)00801741066177(11)240812(17)261220(10)ASHZFC126
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0029YN
Device Lot NumberASHZFC126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2025
Supplement Date Received by Manufacturer04/10/2025
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-