• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 3300TFX
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 02/19/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with a 19mm 3300tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 9 years, 6 months due to stenosis.The patient presented with sob.The procedure was performed with a 20mm 9750tfx transcatheter valve.The patient was discharged pod #13.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections d4 (expiration date), h4, h6 (investigation findings, investigation conclusions).Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.The root cause of this event was determined to be due to patient related factors.The stenosis in this case was impacted by the progression of the patient's underlying valvular disease pathology with or without structural valve deterioration and/or nonstructural dysfunction.The subject device is not available for evaluation as it remains implanted in the patient.The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.H11: corrected data: corrected section h6 (type of investigation).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARD\
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key21600841
Report Number2015691-2025-01898
Device Sequence Number9262402
Product Code DYE
UDI-Device Identifier00690103176124
UDI-Public(01)00690103176124(17)171209
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2017
Device Model Number3300TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2025
Supplement Date Received by Manufacturer04/18/2025
Initial Report FDA Received Date03/13/2025
Supplement Report FDA Received Date04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/10/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age85 YR
Patient SexFemale
-
-