Model Number 100
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
During the shift check, the autopulse platform (sn (b)(6)) displayed the "system error, out of service, revert to manual cpr" message during the device start-up.No patient involvement.
 
Additional Manufacturer Narrative
The reported complaint of the autopulse platform (sn (b)(6)) displayed a "system error, out of service, revert to manual cpr" error message was confirmed during the archive data review and functional testing.The root cause of the system error is related to a communication error that caused a latch-up of the processor board.During the archive data review, system error 132 (internal watchdog timeout) was observed, thus confirming the reported complaint.The autopulse platform failed initial functional testing due to the "system error, out of service, revert to manual cpr" error message displayed upon powering on, thus confirming the reported complaint.Zoll service personnel used the ap vision software to clear the system error.Upon resetting, the platform was re-evaluated through the run_in test using the 95% large resuscitation test fixture (lrtf) with four good known test batteries until discharged without fault or error.Following the service, the autopulse platform passed the final run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries without fault or error.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaints reported for the autopulse platform with sn (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE®RESUSCITATION SYSTEM
Common Device Name
CARDIAC RESUSCITATOR, LINE-POWERED
Manufacturer (Section D)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave.
san jose, CA 95131
MDR Report Key21602023
Report Number3010617000-2025-00200
Device Sequence Number4431212
Product Code DRM
UDI-Device Identifier00849111001014
UDI-Public00849111001014
Combination Product (Y/N)N
Initial Reporter CountryHR
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100
Device Catalogue Number8700-0740-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2025
Initial Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-