| Catalog Number |
62422602 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that an automate iii broke during use.
|
| |
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
03-28-2025: returned product was 1 automate iii tightening device and 1 flexshaft date code 0224 with coil deformation/weld failure causing the product to not function properly.Capa: 2023-519 opened to address automatrix flexshaft assemblies breaking for product manufactured by sarasota since september 2022 (date code 0922) through implementation date of may 2024 (date code 0524).Complaint is considered substantiated.Due to the nature of the case (serious) dhr and retain evaluation has been conducted.
|
| |
|