• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Medical Device Problem Codes Defective Component (2292); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction information provided in h.11.Correction information: the initially submitted mdr report with the ref num (1119421-2025-00683) was consider to be invalid, as it was already submitted with the mfg report number (1119421-2024-02473).All the further information would be reported with the mfg report number (1119421-2024-02473).The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in h.3., h.6.And h.11.The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.A non-qualified viscoelastic was indicated.The root cause for the reported trailing haptic issue my be related to a failure to follow the (instructions for use) ifu.The product was not returned.A non-qualified viscoelastic was indicated.The ifu instructs: during device preparation and implantation of the company iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The ifu instructs: fully insert the ovd cannula, perpendicular to the device, through the viscoelastic port, located in the lens stop portion of the device.Fill the device until ovd can be observed flowing to the fill-to line on the nozzle tip.This will require approximately 0.2 ml of ovd.Only use an company ovd qualified for use with the company pre-loaded delivery system that has been allowed to come to the operating room temperature.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become stuck in the device.Take at least 7 seconds to gently fold the iol by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the fill-to line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be out of position can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported with description faulty lens/injector.Additional information received that the trailing haptic trapped in the injector and tip of injector got snapped.The lens was settled in the bag and was stable.There was no patient harm.The surgeon plans to explant lens.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21604502
Report Number1119421-2025-00683
Device Sequence Number9601896
Product Code KYB
UDI-Device Identifier00380652394826
UDI-Public(01)00380652394826(17)270103(11)240105(10)15776028
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAU00T0
Device Lot Number15776028
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2024
Supplement Date Received by Manufacturer03/14/2025
Initial Report FDA Received Date03/14/2025
Supplement Report FDA Received Date03/14/2025
Date Device Manufactured01/05/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VISCOELASTIC BIOLON X10314AA
Patient Age55 YR
Patient SexFemale
-
-