Model Number ECLC
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that before operation device had no stimulation response.Current stim return showed "0", indicating that current is not being delivered.No patient involved.
 
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa 624392.G4: please note that the involved device is not marketed in the united states; however, it is similar to the united states marketed device (945eclc).H3: product analysis found no fault with the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® CONTROLLER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key21605755
Report Number1045254-2025-00868
Device Sequence Number9456765
Product Code GWF
UDI-Device Identifier00643169378469
UDI-Public00643169378469
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberECLC
Device Catalogue NumberECLC
Device Lot Number212390041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2024
Initial Report FDA Received Date03/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-