• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN AVITUM ITALY S.P.A. APEX¿; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN AVITUM ITALY S.P.A. APEX¿; CONTAINER, I.V. Back to Search Results
Catalog Number 2112532
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: event 1 brief inquiry description 10 3l apex final containers lot 24l04 - black specs inside the chambers - out of box failure - disposable return.Issue(s) reported: stephanie reported that she received 2 packages of 3l final containers where there are black specs inside the containers.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Event 1: a total of ten (10) samples were submitted to the manufacturer for evaluation.Through visual examination, two units with a small dark dot on the spike port were observed.The reported defect was observed.These dots are embedded in the connector material or tubing and are not visible at 50 cm as required by internal procedure for visual inspection.As per the shape and aspect of these dots, they are comparable with burnish material, resulting from the molding process.The presence of these dots does not jeopardize the functionality of the product; the defect is considered aesthetic.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or nonconformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX¿
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola, 41037
IT  41037
Manufacturer (Section G)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola, 41037
IT   41037
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key21606195
Report Number9681240-2025-00008
Device Sequence Number9217761
Product Code KPE
UDI-Device Identifier04046964539220
UDI-Public(01)04046964539220(11)241101(17)271031(10)24L04
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K041415
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2112532
Device Lot Number24L04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/14/2025
Supplement Report FDA Received Date04/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-