EXACTECH, INC. THREE PEG PATELLA 41MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
200-02-41 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Joint effusion (4849)
|
| Date of Event |
08/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: 02-010-01-0360 - lgc femoral ps cem right sz 6; 02-012-44-6013 - logic tibia implant psc insert, sz 6, 13mm; 02-012-45-6060 - lgc tibial fit tray cem sz 6f / 6t; 204-34-02 - fluted stem extension 25l x 14 mm; 204-70-00 - tibial stem ext.Screw.The revision reported was likely the result of the patient¿s reported fall which led to pain.The reported prosthesis wear may have been the result of the orientation of the components, instability, patient related factors, or any combination of these possibilities.However, this cannot be confirmed because the components were not returned for evaluation.Additionally, the insert was packaged in a non-conforming vacuum bag for over five years.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 3.5 years post op initial right tka, this 72 y/o male patient was revised.Patient presented with a sore knee after a fall from a ladder.After opened the knee, surgeon found poly and osteolysis issue.This seems to be from the patella wear.Liner and patella were exchanged.Patient was last known to be in stable condition following the event.No other patient information/medical history reported.Photos & x-rays attached.Product not returning - it was discarded.
|
| |
|
Search Alerts/Recalls
|
|
|