|
|
| Model Number |
HLC-723G8 REFURBISHED |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) was at the customer site to address the reported event.The fse confirmed the reported discrepancy by viewing the chromatograms and noticed hbe suspected and checked peak flags.During troubleshooting, fse found several low and high total area errors, high pressure problems, and rt was at 0.58 minutes.The fse replaced the column which resolved the high-pressure issue.The fse also adjusted the flow rate to resolve the hbe suspected and checked peak flags issue.In addition, the fse replaced the sample needle, small sample syringe, rotor seal, stator face, and rheodyne valve with plate (injection valve assembly) to resolve the low total area issues.Fse repaired and validated the analyzer by successfully performing precision studies without error and within acceptable ranges.No further action is required by field service.The g8 analyzer is functioning as expected.The g8 analyzer, serial number (b)(6) was installed on 06nov2024.A complaint history review and service history review for similar complaints was performed from installation date 06nov2024 through aware date 18feb2025.There were no similar complaints identified during this searched period including this case.The g8 variant analysis mode operator¿s manual v 3.4 under chapter 1: introduction and applications: limitations of the procedure: total area dilution studies demonstrate that the assay is linear from a total area of 500 to 4000.However, the optimum total area is 700 to 3000.Abnormal red cell survival: the life span of red blood cells is shortened in patients with hemolytic anemias, and the actual life span depends upon the severity of the anemia.As a consequence, specimens from such patients may exhibit decreased glycohemoglobin levels compared to patients with normal red cell life span.The life span of red blood cells is lengthened in polycythemia or post-splenectomy patients.Specimens from such patients may exhibit increased glycohemoglobin levels.Hemoglobinopathies: the most commonly observed hemoglobin variants are hbs, hbc, hbd and hbe.These variants in their heterozygous state elute after the hba0 peak in their designated windows: hbs (h-v1 peak), hbd (h-v0 peak) and hbc (h-v2 peak).The hbe (p-hv3 peak) appears between sa1c and hba0 peaks.In all these cases, the sa1c% is reportable when these hemoglobin variants are present in the heterozygous state.When either of these hemoglobin variants is identified, the sa1c% is calculated in a proprietary way, depending on the type of hemoglobin variant, based on the area of sa1c, the area of identified hemoglobin variant and other peaks.When a p-hv3 peak is detected, a flag 43 is also reported.Glycemic monitoring for any patients displaying any homozygous hemoglobin (other than hbaa) such as hbss, hbcc or the double heterozygous sc, cannot be performed using sa1c because there is no hemoglobin a present.Alternative testing is mandatory for these types of patients.The most probable cause of the reported discrepant result was due to failures of the column, sample needle, small sample syringe, rotor seal, stator face, and rheodyne valve with plate (injection valve assembly).
|
| |
|
Event or Problem Description
|
|
A customer reported ¿discrepant patient results for hba1c¿ on the g8 analyzer.The patient sample result of 6.5% was reported out to the physician who questioned the result and requested the sample to be reran.The patient sample was rerun and resulted with 5.2%, which matched the patient history result.The customer also had discrepancies with two other patient samples but did not provide any further information.The column count was around 1500, the analyzer has been calibrated weekly and quality controls (qc) running close to the mean.The customer questioned the retention time on the samples and technical support specialist (tss) recommended the customer use the discrepant patient sample and run a small precision study to check consistency for the results.The customer followed up with tss and stated they performed the precision study and sample results are greater than 5% difference and not within the 5% acceptable criteria for the site.The customer also mentioned receiving hbe suspected flags on all three precisions of 6.5% of the same sample and retention time (rt) 0.58 minutes were within specifications.No patient treatment was affected.A field service engineer (fse) was dispatched to further investigate the reported event.There was no indication of any patient intervention or adverse health consequences due to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record (dhr) was conducted for serial number (b)(6), which confirmed that there were no nonconformance, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the sample syringe and rheodyne valve with plate (injection valve assembly) were returned to tosoh instrument service center (isc) for investigation.A visual inspection was performed on the return parts for cracks, misalignments, corrosion and other physical defects.No damage or defects were observed.Functional testing was performed and all products meet all specifications and is suitable for its intended use.No issues affecting volume accuracy were identified.Isc did not determine the cause of the reported event, cause not established.Correction section h10: after the field service engineer (fse) replaced the column.Sample needle, small sample syringe, rotor seal, stator face, and rheodyne valve with plate (injection valve assembly), the g8 analyzer continued to produce intermittent low and high total area values.Upon further discussion, the customer informed the fse that they did not trust the results from the refurbished analyzer.Due to ongoing issues, the analyzer was replaced and the old analyzer was sent back to tosoh.No further action required by field service.The g8 analyzer is functioning as expected.Additional information for h6: the column and analyzer were added for the component information.
|
| |
|
Search Alerts/Recalls
|
|
|