| Catalog Number |
0036290 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Update 19feb25: the device was returned and evaluated.The completion of the device evaluation confirmed an insufficient heatseal; therefore, this complaint has been reassessed and meets the criteria for a reportable event.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence the distributor reported that the 0036290 device had an insufficient heatseal found during incoming inspection.No patient involvement occurred.The shipment associated with the rga for this device has not yet been returned for evaluation and it has not been confirmed that an insufficient heatseal causing a sterility breach has occurred.Based on the information available, and the decision tree results, this complaint does not meet the definition of a reportable event.Upon the receipt of the device, an evaluation will be conducted and if a breach of sterility is confirmed the reporting decision will be reassessed at that time.
|
| |
|
Additional Manufacturer Narrative
|
|
Received 1 of item 0036290 in unopened original packaging.Performed a visual inspection of the device, there were no obvious signs of a breach.Performed a functional inspection, the device was dye leak tested which indicated that the packaging had insufficient heat seal on the package as there was leakage.A root cause cannot be determined; however, based upon manufacturing process review, a possible cause of this event could be that the issue is related to handling in the transportation of the product.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 10 reports, regarding 11 devices, for this device family and failure mode.(b)(4).We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Search Alerts/Recalls
|
|
|