• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONSOLIDATED MEDICAL EQUIPMENT COMPANY CONMED; TUBE, ASPIRATING, FLEXIBLE, CONNECTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONSOLIDATED MEDICAL EQUIPMENT COMPANY CONMED; TUBE, ASPIRATING, FLEXIBLE, CONNECTING Back to Search Results
Catalog Number 0036290
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Update 19feb25: the device was returned and evaluated.The completion of the device evaluation confirmed an insufficient heatseal; therefore, this complaint has been reassessed and meets the criteria for a reportable event.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence the distributor reported that the 0036290 device had an insufficient heatseal found during incoming inspection.No patient involvement occurred.The shipment associated with the rga for this device has not yet been returned for evaluation and it has not been confirmed that an insufficient heatseal causing a sterility breach has occurred.Based on the information available, and the decision tree results, this complaint does not meet the definition of a reportable event.Upon the receipt of the device, an evaluation will be conducted and if a breach of sterility is confirmed the reporting decision will be reassessed at that time.
 
Additional Manufacturer Narrative
Received 1 of item 0036290 in unopened original packaging.Performed a visual inspection of the device, there were no obvious signs of a breach.Performed a functional inspection, the device was dye leak tested which indicated that the packaging had insufficient heat seal on the package as there was leakage.A root cause cannot be determined; however, based upon manufacturing process review, a possible cause of this event could be that the issue is related to handling in the transportation of the product.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 10 reports, regarding 11 devices, for this device family and failure mode.(b)(4).We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONMED
Common Device Name
TUBE, ASPIRATING, FLEXIBLE, CONNECTING
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX  31136
Manufacturer (Section G)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX   31136
Manufacturer Contact
robin drum
11311 concept blvd
largo, FL 33773
8653881978
MDR Report Key21617293
Report Number3007305485-2025-00046
Device Sequence Number12446033
Product Code BYY
UDI-Device Identifier10653405000471
UDI-Public(01)10653405000471(17)280103(10)202301046
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number0036290
Device Lot Number202301046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2025
Initial Report FDA Received Date03/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-