• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DS2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DS2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number FRX521H14C
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging that the ds2adv auto cpap device does not light up.There was no report of serious patient harm or injury.There was no report that patient required medical intervention.The device was returned to a third-party service center for evaluation and the reported customer complaint was reproduced.During evaluation of the device, the third-party service center visually inspected the device and found no light from display, ui bezel defective, main board burn, blower outlet seal deformed, blower contaminated inside, in/out seal missing and new pollen filter must added.Lastly, the device was scrapped as per customer request.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DS2ADV AUTO CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key21617582
Report Number2518422-2025-035051
Device Sequence Number6837703
Product Code BZD
UDI-Device Identifier00606959062757
UDI-Public(01)00606959062757
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K200480
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberFRX521H14C
Device Catalogue NumberFRX521H14C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2024
Initial Report FDA Received Date03/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-