• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I GOLD-TITE SQUARE UNISCREW; DENTAL SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I GOLD-TITE SQUARE UNISCREW; DENTAL SCREW Back to Search Results
Catalog Number UNISG
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie (b)(4).A2: age and date of birth unknown / not provided.A3: gender unknown / not provided.A4: patient weight unknown / not provided.B3: date of event unknown / not provided.E1: last name unknown / not provided.
 
Event or Problem Description
It was reported that screw loosening and was retightened.
 
Additional Manufacturer Narrative
Zimvie did not receive one (1) unisg, (gold-tite square uniscrew) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Dhr review was completed for the subject lot number 1200249.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1200249 for similar events and one other complaint was identified ((b)(4)).Review completed utilizing keywords: ¿dental : functional : loosening.¿ the customer did not submit images for the reported event.Based on the investigation and risk management file review as per rm-00057-haz rev 22, the most likely root cause determined from the investigation was abutment screw is not torqued to specification/probably user caused.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOLD-TITE SQUARE UNISCREW
Common Device Name
DENTAL SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21617916
Report Number0001038806-2025-00583
Device Sequence Number8510619
Product Code NHA
UDI-Device Identifier00844868017325
UDI-Public(01)00844868017325(17)211011(10)1200249
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K072642
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/11/2021
Device Catalogue NumberUNISG
Device Lot Number1200249
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/17/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/11/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-