| Model Number |
K062 |
| Medical Device Problem Code |
Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker device exhibited undersensing.Sensing fixed was 2.5 milli volts output.As of this time, this device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker device exhibited undersensing.Sensing fixed was 2.5 milli volts output.As of this time, this device remains in service.No adverse patient effects were reported.Additional information was received which indicated that this pacemaker was explanted, replaced and was returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker device exhibited undersensing.Sensing fixed was 2.5 milli volts output.As of this time, this device remains in service.No adverse patient effects were reported.Additional information was received which indicated that this pacemaker was explanted, replaced and was returned for analysis.No additional adverse patient effects were reported.
|
| |
|