MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25; HYDROCEPHLALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25; HYDROCEPHLALUS MANAGEMENT Back to Search Results
Model Number FX643T
Medical Device Problem Codes Infusion or Flow Problem (2964); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 01/28/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav 2.0 shunt system (#fx643t) was implanted on an unspecified date.According to the complainant, the shunt sytem caused an overdrainage and had adjustment difficulties.The patient underwent a revision procedure on (b)(6) 2025.No patient complications were reported as a result of the revision procedure.The complained product was sent to the manufacturer for investigation.
 
Additional Manufacturer Narrative
Visual inspection: during the investigation, scratches on the outer housing of the valves, but no significant deformations or damage were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves do not operate within the permissible tolerance in their respective relevant positions an accelerated outflow of both valves could be determined.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected; however, the breaking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in both valves.Results: based on our investigation results, we can determine an accelerated outflow of both valves and adjustment difficulties in the progav 2.0.The determined deposits can be named as the cause for the functional deviations.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25
Common Device Name
HYDROCEPHLALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key21618100
Report Number3004721439-2025-00075
Device Sequence Number8516154
Product Code JXG
UDI-Device Identifier04041906502010
UDI-Public4041906502010
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/02/2023
Device Model NumberFX643T
Device Catalogue NumberFX643T
Device Lot Number20037339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2025
Initial Report FDA Received Date03/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexMale
Patient Weight14 KG
-
-