SYNTHES GMBH TFNA FENESTRATED SCREW 95MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
04.038.195S |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon encountered abrasion when inserting the femoral neck screw.There was a problem with advancing the screw - it touched the lateral part of the nail.It couldn't be inserted securely and safe through the lateral nail opening.Another implant (nail and screw) were opened.Fragments generated were easily removed without additional intervention.Surgery was completed successfully with a delay of unknown length.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10.H3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the tfna fenestrated screw 95mm - sterile was observed stripped around the proximal thread area.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.A dimensional inspection for the tfna fenestrated screw 95mm - sterile was not performed due to post manufacturing damage.A functional test was performed connecting the mating tfna nail to the tfna screw; there was no problem with advancing the tfna screw, the tfna screw passed freely through the nail hole, and the locking mechanism in the nail advance accordingly.The tfna screw could be inserted securely and safely through the nail opening.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the tfna fenestrated screw 95mm - sterile would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: manufacturing location: elmira ¿ packaged, labeled, sterilized and released by: monument.Manufacturing date: 12-dec-2024.Expiration date: 01-nov-2034.Part number: 04.038.195s, tfna fenestrated screw 95mm -sterile.Lot number: 45376p4 (sterile).Device history review: a manufacturing record evaluation was performed for the finished device with batch number 45376p4.No nonconformities or manufacturing irregularities have identified related to the complaint condition.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|