|
The reported event was confirmed ¿ cause unknown.The reported failure was able to be reproduced.The product was used for urological care.Evaluation confirmed the balloon was burst.Introduced water into the inflation lumen and did not experience any obstructions to impede flow.Microscopic examination found there was present salt on the balloon area.However, the exact cause of how and when the problem occurred could not be determined.The potential root cause for this failure could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/operator error).A review of the device history record did not show any problems or condition that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "warnings 1.Method for use do not inflate the balloon in the urethra.The urethra may be injured.Do not pull the catheter hard.The bladder/urethra may be injured.2.Applicable patients (1) patients with delirium who might pull out catheter when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.Contraindications 1.Method for use: do not reuse.Do not resterilize.Be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).They may damage the device and may burst balloon.Do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.Catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.2.Applicable patients (1) do not use in patients who are or have been allergic to natural rubber latex shape, configuration and principles bardex® biocath® foley catheter is a balloon catheter.There are different type variations, for example, hematuria types with a catheter internally reinforced with nylon fiber for preventing occlusion of its inner lumen and securing urinary passage.Some may include a water-soluble lubricant as an accessory.Material balloon catheter: natural rubber latex sizes of catheters available in sizes 8 to 26 every 2fr8fr and 10fr: provided with white straightener 2.Accessories water-soluble lubricant (may not be included in some product variations.) intended use & effect- efficacy the device is designed to be placed in the bladder for the purpose of urinary drainage, bladder irrigation or hemostasis.Directions for use 1.Method of use (1)the device is intended for single use only and is not reusable.(2)cleanse the area around the external urethral meatus with the cotton balls immersed in the antiseptics.(3)lubricate the distal end of the catheter with water-soluble lubricant.Insert catheter into the urethral meatus and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a needle-less syringe, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.As to double-balloon catheter, infuse through the valve printed bladder for balloon inflation.After the prostatic balloon part of the device enters the prostate, using a needleless syringe, infuse the specified volume of sterile water through the valve printed prostate to inflate the balloon.(4) pull catheter to seat the balloon at the level of the bladder neck and secure placement.(5)to deflate balloon and remove catheter, insert a luer tip (needle-less) syringe in the inflation valve to allow the water to drain spontaneously.After balloon deflation, withdraw the catheter while confirming that no resistance is encountered.2.Precautions for use (1)when resistance is encountered in inserting catheter, stop the procedure and remove the catheter.(2)when deflating balloon, do not aspirate with a syringe.The inflation lumen for balloon deflation may be occluded by negative pressure, and as a result the catheter cannot be removed.(3)when inserting catheter with a stylet, confirm that the stylet has reached the tip of the catheter, and make sure to keep the stylet in place inside the catheter during insertion.(4)for catheters with a white straightener, first remove the straightener before lubricating the catheter as inserting the catheter with the straightener may cause urethral injury.(5)as to double-balloon catheter, the valve is labeled "20/40ml/cc"; however, infuse 25 ml of sterile water into the cap printed bladder and 45ml into the cap (6) no substance except sterile water should be used to inflate the balloon.(7) do not wipe catheter surface with organic solvents such as alcohol.(8do not aspirate urine through drainage funnel wall.(9)since movement of the body, etc.May twist or bend catheter to cause occlusion, care should be taken to fix the catheter securely.(10)when urinary flow cannot be noted, confirm that the catheter is neither occluded nor bro-ken.(11)when irrigating, open the cap of irrigation funnel and attach irrigation line or tube under aseptic technique.After use, close the cap.Precautions 1.Precautions for use (exercise caution when using the device in the following patients) (1)exercise caution when using the device in patients with high urinary calcium levels as encrustation on the balloon surface, catheter occlusion or damage may occur.2.Important precautions (1)when catheter is inadvertently removed, inspect the balloon and shaft of catheter for rupture, defect, etc.Before inserting a new catheter.(2)when any part of the balloon and/or the catheter is missing, consider removing them using a cystoscope.(3)when it is difficult to remove catheter by deflating the balloon, take appropriate measures according to the section troubleshooting when it is difficult to remove catheter by deflating the balloon (expressed as "removal-difficult case" hereinafter), take appropriate measures according to the following procedures.The following two methods are available for removal-difficult cases a.Non-rupture method (sterile water is withdrawn without bursting the balloon.) b.Balloon-rupture method with balloon-rupture method, fragments of the ruptured balloon may remain in the bladder.Therefore, try non-rupture method first.A.Non-rupture method 1) attach luer tip syringe to the inflation valve.Inject an additional amount of sterile water into the inflation lumen and pump the plunger.2) if situation wouldn't be improved with 1), sever the inflation funnel of valve.(fig.1) 3) if situation wouldn't be improved with 2), sever the catheter shaft while holding it with for-ceps so that the distal segment may not be drawn into the urethra.(fig.2)fig.2 4) if situation wouldn't be improved with 3), insert a needle into the inflation lumen and pump the plunger.(fig.3)fig.3 5) if situation wouldn't be improved with 4), insert a fine steel wire through the inflation lumen of catheter.(fig.4)fig.4 b.Balloon-rupture method 1) inject 100-200ml/cc of saline solution warmed to body temperature into the bladder through the drainage lumen, and then inject a large amount of water or 10-15 ml/cc of mineral oil into the balloon through the inflation lumen with a needle to induce rupture.After rupture of the balloon, irrigate the bladder.(fig.5) 2) if situation wouldn't be improved with 1), attempt following procedures.A) under the radioscopic observation, infuse a contrast medium into the bladder, and burst the balloon by suprapubic puncture of the bladder.(fig.6) b) in male patients, burst the balloon by puncture with a needle from the perineal (or suprapubic) region or through the rectum under ultrasonographic guidance.(fig.7) c) in female patients, burst the balloon by insertion of a needle along the urethra.(fig.8) 3.Malfunction and adverse events (1) malfunction -catheter kinking, damage, rupture -difficulty or failure to remove the device -occlusion of catheter inner lumens -encrustation -accidental removal of the device due to leakage of sterile water or balloon rupture -device damage due to inappropriate use (2) adverse events -urinary-tract infection -hemorrhage, hematuria -allergy reaction to the device -calculus formation -edema -pain -discomfort -injury of bladder or urethral -urethritis, urinary incontinence -retained balloon fragments storage method and expiration date storage store in a dry, cool place away from heat, moisture, and direct sunlight.Expiration date indicated on the direct package and the outer box." this report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|