• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O. HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TURIS; WA22706S

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O. HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TURIS; WA22706S Back to Search Results
Model Number WA22706S
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the plasma loop handles were burned.There were no reports of patient harm.
 
Event or Problem Description
There was rupture of the loop and interruption of the electric current.The first loop burned soon after the first pedal activation already inside the patient's urethra.The second one the physician tested prior to use before contacting the patient (outside patient), where it burned and got dark with a burning smell.The procedure time increased about 15 minutes, due to the necessity of changing the device plus the equipment.The event occurred during a prostate hyperplasia procedure.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation, and update to b5, d4, h4, and h6 component code.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TURIS
Common Device Name
WA22706S
Manufacturer (Section D)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ  75124
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ   75124
Manufacturer Contact
todd brill
801 west park drive
westborough, MA 01581
5082077661
MDR Report Key21619463
Report Number3007960282-2025-00034
Device Sequence Number15564888
Product Code FAS
UDI-Device Identifier14042761085356
UDI-Public14042761085356
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K171965
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA22706S
Device Lot Number1000150600
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/21/2025
Supplement Date Received by Manufacturer04/25/2025
Initial Report FDA Received Date03/17/2025
Supplement Report FDA Received Date04/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WA22706S.
Patient SexUnknown
-
-