BOSTON SCIENTIFIC CORPORATION XXL ESOPHAGEAL; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M001145540
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon got punctured by the overlapping stents and additional intervention required to remove the balloon material.The 40% stenosed target lesion was located in the common femoral vein.A 16-6/5.8/75 xxl balloon catheter was selected to post-dilate and place two wallstents for venous compression.During the procedure, at the overlapping segment of the two stents, the balloon catheter was punctured.As the physician attempted to remove the damaged balloon; however, it broke and became lodged in the common femoral vein.Despite efforts to retrieve the balloon percutaneously, removal through the sheath was unsuccessful.Given the circumstances, no further minimally invasive options were feasible, necessitating a surgical cut-down procedure to extract the fractured balloon.The procedure was successfully performed, and the incision site was closed without complications.The patient was admitted to the hospital for observation but is anticipated to recover fully without any further issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XXL ESOPHAGEAL
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21620863
Report Number2124215-2025-14344
Device Sequence Number8515745
Product Code KNQ
UDI-Device Identifier08714729163114
UDI-Public08714729163114
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2026
Device Model NumberM001145540
Device Catalogue NumberM001145540
Device Lot Number0032353476
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2025
Initial Report FDA Received Date03/17/2025
Date Device Manufactured09/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age85 YR
Patient SexMale
-
-