OSCOR INC. OCCLUTECH DELIVERY SET III; INTRODUCER, CATHETER Back to Search Results
Model Number 98DS010
Medical Device Problem Codes Fluid/Blood Leak (1250); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Air Embolism (1697)
Date of Event 03/05/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Prior to implantation occluder was flushed, loaded 2x and always flushed with water to remove air bubbles from the device and material.Sheath was flushed as well and once placed in the patient we aspirated 20ml of blood to remove air from the sheath after dilator removal.Then loader together with the occluder was under constant flush connected to the delivery sheath.After the left side of the occluder was deployed ecg has changed and the patient felt chest pain and shortness of breath.Air embolie was suspected.After 5-7 min ecg was again normal and the chest pain was less present in the patient.The device was implanted, and the patient was discharged without any consequences.Device was discarded by hospital.Consequences to event: air entering into vascular system & embolism.(b)(6)2025: customer provided additional details: procedure taking place when issue was found was to close the pfo (patented foramen ovale).It is unknown exactly when the air got into the device but it was noticed (ecg changed and patient complained about the chest pain and breathing issues) shortly after the left side of the pfo device was deployed.There was a short delay of 10-15 more minutes until the air was flushed from the heart and patient started feeling better.The patient received sublingual nitroglycerin and got better.We had to wait until the air was flushed out of the heart and as mentioned above second patient received nitroglycerin just to ease her pain/air embolism.Procedure was completed successfully.
 
Additional Manufacturer Narrative
Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Additional Manufacturer Narrative
The following sections were updated in follow-up 1.B4, g3, g6, h2, h6 & h11.No product was provided for analysis as product was discarded by customer.The customer stated the sheath was flushed then aspirated with 20ml of blood to remove air from the sheath.Ecg has changed and the patient felt chest pain and shortness of breath.Air embolie was suspected.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.No product was returned to oscor inc.For evaluation; however, a complaint notification was provided.Without the return of the actual device in question for evaluation, oscor inc.Is unable to determine the exact cause for this incident.At this time, it is not possible to assign a definitive root cause for the event as reported.The complaint is non-verifiable as the product was not returned for evaluation.Ifu: potential adverse effects include air embolism.Based on the investigation, a capa is not required.In addition, there was no new failure mode identified, and the risk remains acceptable.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may have not been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCCLUTECH DELIVERY SET III
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR INC.
4875 palm harbor blvd.
palm harbor FL 24683
Manufacturer (Section G)
OSCOR INC.
4875 palm harbor blvd.
palm harbor FL 24683
Manufacturer Contact
ronnie greco
4875 palm harbor blvd.
palm harbor, FL 24683
7279372511
MDR Report Key21621822
Report Number1035166-2025-00010
Device Sequence Number12446066
Product Code DYB
UDI-Device Identifier00885672010300
UDI-Public00885672010300
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K210627
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number98DS010
Device Catalogue Number98DS010
Device Lot NumberDP-21549
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/07/2025
Supplement Date Received by Manufacturer04/11/2025
Initial Report FDA Received Date03/17/2025
Supplement Report FDA Received Date04/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening;