• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. INDEFLATOR; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. INDEFLATOR; ACCESSORIES Back to Search Results
Model Number 1000184
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.It is possible that the connection port was not fully connected/tightened and/or the device being connected was compromised thus resulting in the reported leak difficulties; however, as the device was not returned for analysis this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported leak difficulties.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure performed was to treat a de novo, heavily calcified, heavily tortuous right coronary artery.The 20/30 indeflator pressure would not increase during the procedure.An unspecified device was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDEFLATOR
Common Device Name
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key21624199
Report Number2024168-2025-02750
Device Sequence Number17849597
Product Code MAV
UDI-Device Identifier08717648013591
UDI-Public(01)08717648013591(17)250228(10)60440894
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K991102
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number1000184
Device Catalogue Number1000184
Device Lot Number60440894
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2025
Initial Report FDA Received Date03/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-