Model Number 0684-00-0568-01U
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during intra-aortic balloon therapy, the pump showed "sensor failure" error message.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id : (b)(4).
 
Additional Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h6 (type of investigation , investigation findings, component codes, investigation conclusions), h11.Corrected fields: b5, h6 (health effect ¿ clinical code, health effect ¿ impact codes).No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Complaint id: (b)(4).
 
Event or Problem Description
Dr.(b)(6) called about the message ¿sensor failure¿ on the pump.It was pumping and the patient is stable.No report of harm or injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 7.5FR. 40CC IAB WITH ACCESSORIES, US ONLY
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer Contact
henry pena
15 law drive
fairfield, NJ 07004
MDR Report Key21628890
Report Number2248146-2025-0000148
Device Sequence Number18197702
Product Code DSP
UDI-Device Identifier10607567109565
UDI-Public(01)10607567109565(10)3000315480(17)260608(11)230608
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K122628
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2026
Device Model Number0684-00-0568-01U
Device Catalogue Number0684-00-0568-01U
Device Lot Number3000315480
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2025
Supplement Date Received by Manufacturer04/29/2025
Initial Report FDA Received Date03/18/2025
Supplement Report FDA Received Date04/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
IABP - UNKNOWN
Patient Age67 YR
Patient SexMale
Patient Weight116 KG
-
-