| Model Number |
VLFT10GEN |
| Medical Device Problem Codes |
Application Program Problem: Parameter Calculation Error (1449); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: lf1823, blunt tip open sealer divider lf1823 (lot#:32490477x) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.An evaluation was led of a vlft10gen for a non-reportable condition involving an error message with a ligasure device.During the investigation a generator software issue was identified as being related to the reported error message in which this error message may occur when the vlft10gen generator performs an authentication check to confirm that the ligasure instrument has not been used previously.It was reported that the device had an error code.The reported issue was confirmed.The most likely cause was traced to inadequate software design.This identified software issue has a potential for patient harm.Software enhancements have been implemented to mitigate this condition.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter during lysis of adhesions, the device had error code 420 maximum number of uses reached immediately after connection while plugging the device into the generator.Another ligasure device was used successfully with this generator.The device was not reprocessed by another company and not re-used on a patient.There was no patient injury.Medtronic's initial evaluation of the incident device found an ft10 generator software issue related to the customer's operating software version in which the authentication check gives an incorrect error message.
|
| |
|
Search Alerts/Recalls
|