Catalog Number 297529085
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that after 85mm body and 40mm neck had been torqued together, the neck still spun on the proximal body trial.This made dislocating the hip after trialing difficult.There was 2 minutes of surgical delay.There was no patient consequences.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary according to the information received, " it was reported that after 85mm body and 40mm neck had been torqued together, the neck still spun on the proximal body trial.This made dislocating the hip after trialing difficult." the product was not returned to medtech orthopaedics; however, photos were provided for review.¿(b)(4) device photo ad (b)(6) 2025 and (b)(4) device photo ad (b)(6) 2025 - 1.¿ the photo investigation did not reveal that reclaim prox trial shaft 85 (297529085) had any assembly or lock-unlock issue.The evidence provided was not sufficient to confirm the reported event.Functionality issues cannot be evaluated through photo investigation.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the reclaim prox trial shaft 85 would not contribute to the complaint about the device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received stating, "no surgical delay.There were no patient consequence.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM PROX TRIAL SHAFT 85
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21630634
Report Number1818910-2025-04097
Device Sequence Number8350126
Product Code HWT
UDI-Device Identifier10603295156277
UDI-Public(01)10603295156277(10)SO2049176
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number297529085
Device Lot NumberSO2049176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2025
Supplement Date Received by Manufacturer03/28/2025
05/08/2025
Initial Report FDA Received Date03/18/2025
Supplement Report FDA Received Date03/31/2025
05/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/30/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age39 YR
Patient SexFemale
-
-