| Catalog Number |
130762138 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Device history batch: null.Device history review: null.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent revision surgery due to dislocation - some osteophytes but the surgeon commented he should have lateralised more as per materialise plan.Surgeon lateralised the glenosphere an extra 2mm (originally 2mm and lateralised to 4mm) also replaced +3/38mm poly to +6 retentive / 38mm.There was nothing wrong with the products; the patient's anatomy of the scapular neck was impinging on the original glenosphere so by lateralising and upsizing the cup, there was no impingement.The patient had dislocated 3 times, with mua the first two times and revision surgery the third time.The shoulder was dislocating due to having ostephytes present still as well as not being lateral enough, and the implants were impinging on the neck of the humerus.Affected side: left.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Search Alerts/Recalls
|