|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d5, d10, g3, g6, h1, h2, h3, h4, h6, h11 visual inspection of the returned device confirms that several teeth near the tip are excessively worn out.The device also shows wear consistent with usage.Review of the device history records found no related deviations or anomalies.Reported event is not related to a combination of products; therefore a compatibility review is not applicable.Medical records were not provided.This complaint is confirmed via device evaluation.Root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|