Titan pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.A separation was noted on the exhaust tubing of cylinder 1; this is a site of leakage.The surfaces appear to have crease marks adjacent to or within a confined area of abrasion, indicating the area was kinked and abrading against itself for an extended period of time.A separation was noted on the bladder of cylinder 2.This is a site of leakage.The separation had surfaces with a darkened or melted appearance, indicating contact with a cauterizing tool.Abrasion was noted on the inlet tube of the reservoir.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that the abrasion marks noted on all pump tubing matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo.This then may have caused the exhaust tube of cylinder 1 to be kinked onto itself for a period of time.This positioning, in combination with device usage over time, could contribute to sufficient stress to separate the exhaust tubing.A separation of this type could then allow the loss of fluid, making the device inoperable.Because these components were released according to manufacturing and quality control procedures, it was concluded that the observed instrument separation in the bladder of cylinder 2 occurred after the device packaging was opened.In addition, because the expected use of this device combined with the observed separation would have resulted in an earlier detected fluid loss, it was concluded that the separation most likely occurred during or after explant.This separation is not associated with the cause for failure.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|