| Model Number |
DRIVER SHAFT, T-15, SHORT |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 03/03/2025, a sales representative reported via sems-(b)(4) that a (qty.(b)(4)) of an ar-9545-t15-01 driver shaft and an ar-9545-t15-03 driver shaft was twisted and bent over time and unable to be used as intended.The case was completed without adverse effects on the patient.This was discovered during a procedure, with no reported patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-9545-t15-01 batch 8001845 was received for investigation.Visual evaluation identified damage to the hex geometry.The driver's tip had nicks and was twisted.The laser marks are faded and corroded.Functional testing cannot be performed due to the device's damage.The most likely cause of the reported failure is wear and tear from repetitive usage.The device was manufactured on 27 november 2018.
|
| |
|
Search Alerts/Recalls
|