|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H6 component code: g07002 device not returned.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The single complaint was reported with multiple events.There are no additional details regarding the additional patient events.Attempts were made to obtain the information.If further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: - please confirm the number of patients affected by this alleged deficiency (".Over the course of several weeks during total vaginal hysterectomys, specially during the vaginal cuff closure.While using a vicryl suture the suture broke away from the needle.")? - i do not have the number of patients at this time.- have any of these events been previously reported to ethicon? if so, provide the respective reference number(s).- they have not, this is the first reported event.Please provide information requested below for each patient/event.- were there patient consequences? if yes, please explain.- no patient consequences to report.- can you identify the lot numbers of product codes vcp215h and vcp417h? - i am not able to identify the lot numbers from the suture that he reported to me had broken away from the needle.- please confirm below, how many devices demonstrated the alleged deficiency? j260h - lot number 1046xj: - unknown.Vcp215h - lot number unknown: - unknown.Vcp417h- lot number unknown: - unknown.- please confirm if the devices are available for product return.If yes, confirm the quantity of devices available for product analysis and provide the status of the devices as they have not been received for analysis.If the device has been shipped, please provide the shipment tracking details.- product has been shipped and confirmed to have been received - provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep).- dr.(please refer the attached email- re: 2nd product complaint follow-up (b)(4)/suture/r) - please provide additional details of the reported alleged deficiency ("¿two needles, codes unknown are also available that were retained.").- please confirm, did the needles fall into the patient(s)? if not, please explain.No, they did not.If yes, please provide the following information: - were there patient consequences? if yes, please explain.- were x-rays taken to locate the needles? - were the needles removed from the patient(s) during the same procedure? - was there any additional tissue damage as a result of searching for the needles? *if not retrieved, in what tissue structure the needle piece(s) were retained? is there any plan in place to remove the needle(s) in the future? please provide details.- if available, please provide the product codes and lot numbers.Just wanted to confirm that these are just additional users (as the team was wanting to understand if other surgeons were having pull off issues) and there hasn¿t been any new complaints that need to be created for these codes or these surgeons? - these are just additional users of these sutures.We discussed on our call finding out if there were other surgeons in the hospital using these exact codes.The or manager provided me a list of names that may be helpful with discussion with dr.(b)(6) as there has been no complaints of suture detaching from needles from those individuals listed.In addition, is vcp311h a new complaint from dr.(b)(6) or a new complaint from another surgeon? - vcp311h should have been made part of the initial complaint.I did call this code in today and they are generating a new complaint number for this but noting that it was sent in with the additional boxes sent to me for (b)(4).This box has already been shipped from stl the product vcp311h was not in the original product complaint, but i was sent to you with the action of the (b)(4).It has been returned already.
|
|
It was reported that a patient underwent a total vaginal hysterectomy on an unknown date and suture was used.During the procedure, upon vaginal cuff closure, the suture broke away from the needle.When applying tension after the surgeon takes a bite, when they grasp the suture is popping off.There were no adverse patient consequences reported.Additional information was requested.
|