| Model Number |
19-0126 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product has returned but the investigation is not complete at this time.When the findings become available a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a screw was backing out of the cetra plate in a revision case on (b)(6) 2025.The surgery was for an adjacent level and it was reported that the plate was being removed anyway.The patient was not experiencing any symptoms.It was reported that the locking plate was intact when the device was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
The 1-level plate assembly (pn: 19-0126 ln: e19) was returned for investigation.The plate was returned with 4 screws.Radiographic images were provided to assist with the investigation and confirm that one screw had backed of the construct.The edges of the locking plate that broke off have deformation, likely from contacting the head of the screw.The screw heads also show deformation.The locking tab likely failed because of ductile overload and ductile shear overload on the screw.The locking tab was likely subjected to overload by leaving the screws proud causing the locking tab to stress at fillet which is weaker cross section of the design.A review of the product dhr showed that all inspections met design specifications.A review of the ncmr databases did not reveal any nonconformances related to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that a screw was backing out of the cetra plate in a revision case on (b)(6) 2025.The surgery was for an adjacent level and it was reported that the plate was being removed anyway.The patient was not experiencing any symptoms.It was reported that the locking plate was intact when the device was removed.
|
| |
|