| Catalog Number |
386865 |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Our quality engineer inspected the photo submitted for evaluation.The reported issue of safety shield activation failure was confirmed upon inspection of the photo.However, bd cannot confirm the cause of the failure to our manufacturing process since no physical sample was returned.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd us cathena 18gx1.25in straight bc safety shield activation failed.It was reported by customer that cathena needle not safetying.Verbatim: cathena needle not safetying.
|
| |
|
Search Alerts/Recalls
|