|
The lens case lid for this serial number was returned.A broken haptic was adhered to the underneath of the lens case lid.The other loose half optic appeared to go to this file.Two used company cartridges were also returned.It is unknown which cartridge was used with each lens.Evaluation was placed in both files.The company cartridge was microscopically examined.Viscoelastic was observed in the cartridges.Both cartridge tips had heavy stress.One had and an aneurysm on the left side of the tip.The second cartridge tip had two aneurysms, one on the left side of the tip and one bottom center.Both cartridges had evidence of placement into a handpiece.The company cartridges were cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Photos were provided.In both photos, a broken haptic was visible under the optic.This may indicate the haptic was misfolded under the optic instead of folded onto the anterior surface.No other determination can be made from the provided photo.Product history records were reviewed and documentation indicated the product met release criteria.Qualified associated products were indicated.The root cause for the broken haptic could not be determined.Based the damage observed to the returned cartridges, the lens or plunger may not have been in acceptable positions for advancement.Both cartridges had heavy stress and aneurysms.Top coat dye stain testing was conducted with acceptable results.The photos showed the broken haptic under the lens.This may indicate the haptic was misfolded under the optic.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|