MAUDE Adverse Event Report: B BRAUN MELSUNGEN AG EASYPUMP 2; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B BRAUN MELSUNGEN AG EASYPUMP 2; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 4540016-07
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
According to the event description: the easypump ran through in 22 hours instead of 50 hours.No patient complications were reported as a result of this event.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun internal report number (b)(4).Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.Device history record (dhr): reviewed the dhr for batch 23h28ged81, there was no defect encountered during in process and final control inspection.No sample and no picture received for further investigation.Investigation results: final control flow rate report of affected batch 23h28ged81 was reviewed.For final control flow rate report, the average flow rate deviation from the nominal flow rate was between -4.97% and 3.17%.As no complaint sample was received, further investigation was not possible.There are some possibilities that can cause the sample to empty faster than nominal time in actual application, such as one or combination factors of more than one as below: factor 1: temperature the temperature of the surrounding will affect the flow rate of the sample.For every increase of 1°c, the flow rate of the sample will increase approximately by 3%.For example, if the flow restrictor of the product reaches the temperature of 37°c, the flow rate of the sample will increase by approximately 18% which means the flow rate will increase from 2 ml/hr to 2.36 ml/hr.Factor 2: folded outer shell (outer layer) folded outer shell (outer layer) will also act as the external pressure to elastomeric membrane and increase the flow rate of the product.Factor 3: external pressure external pressure such as squeezing or laying on the pump will increase the flow rate which cause the pump to empty earlier than the nominal time.Simulation of external pressure had been conducted.The flow rate of the product can increase when external force is applied to the pump.However, this external force is against the intended use of the product according to ifu.Conclusion: as no complaint sample was received, further investigation was not possible.Therefore, this complaint is considered as not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP 2
Common Device Name
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
B BRAUN MELSUNGEN AG
carl-braun str. 1
melsungen, hessen 34212
GM  34212
Manufacturer (Section G)
B BRAUN MELSUNGEN AG
carl-braun str. 1
melsungen, hessen 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key21634982
Report Number9610825-2025-00133
Device Sequence Number9372148
Product Code MEB
UDI-Device Identifier04046964448607
UDI-Public(01)04046964448607(11)230828(17)280801(10)23H28GED81
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4540016-07
Device Lot Number23H28GED81
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2025
Initial Report FDA Received Date03/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-