Note this report is for the malfunction of sample 1.A subsequent mdr will be filed for sample 2.Investigation summary: two (2) photos were provided by the customer, unable to confirm complaint from the photos provided.Receive done (1) used list# b33982, 5.5" (14 cm) appx 1.2 ml, smallbore quadfuse ext set w/3 clave¿, 3 check valves, 4 clamps (red, white, green, yellow), luer lock, one (1) used list# unknown, 2-way extension set with catheter.One (1) used list# b33982, 5.5" (14 cm) appx 1.2 ml, smallbore quadfuse ext set w/3 clave¿, 3 check valves, 4 clamps (red, white, green, yellow), luer lock.As received, one of the claves on sample #1 was stuck down, no other anomalies observed both samples were leak tested, confirmed leaking on both samples, after disassembly, one returned sample (sample#2) showed signs of insufficient solvent, sample #1 was retuned with one clave stuck down, after disassembly, the spike showed signs of being mated with an incompatible mating device.Complaint of leaking is confirmed, probable cause of leaking for sample #1 is due to incompatible mating device.Probable cause of leaking for sample#2 is due to insufficient solvent during assembly.
|
|
A complaint was received regarding a 5.5" (14 cm) appx 1.2 ml, smallbore quadfuse ext set w/3 clave¿, 3 check valves, 4 clamps (red, white, green, yellow), luer lock that experienced leakage.This is 1 of 2 reports.The report stated that the device is leaking.Sample photos are provided.There was unknown patient involvement and unknown patient harm.
|